This study was a pilot study conducted at the 108 Government Military Hospital in Hanoi, Vietnam, whereby randomly selected patients regardless of body weight were sent for whole body 18F-FDG PET imaging with a dose of mCi/kg, which is the lower threshold in the recommended dose range in the guidelines on the Society of Nuclear Medicine and Molecular Imaging website. | Journal of Science & Technology 128 (2018) 032-035 PET Image Quality Evaluation at Upper and Lower Threshold of 18F-FDG dose: A Pilot Study in Vietnamese Patients Nguyen Thai Ha1*, Duong Trong Luong1, Nguyen Thanh Trung2, Nguyen Thu Van1 1 Hanoi University of Science and Technology, Hanoi, Vietnam 2 108 Government Military Hospital, Hanoi, Vietnam Received: March 28, 2018; Accepted: June 29, 2018 Abstract 18 F-FDG Positron Emission Tomography (PET) is the most common modality used in cancer diagnosis and monitoring. PET imaging results in patients being exposed to a certain level of radiation, which may be harmful to the body. Research has been conducted on reducing radiation dose in PET imaging in order to minimize radiation risks, at the same time still maintaining diagnostic image quality. This study was a pilot study conducted at the 108 Government Military Hospital in Hanoi, Vietnam, whereby randomly selected patients regardless of body weight were sent for whole body 18F-FDG PET imaging with a dose of mCi/kg, which is the lower threshold in the recommended dose range in the guidelines on the Society of Nuclear Medicine and Molecular Imaging website. Another group of patients underwent the same PET procedure with dose calculation using the upper threshold of the same range ( mCi/kg). Image quality, manifested through the Signal-to-Noise Ratio (SNR), was compared between the low-dose and high-dose groups. No statistically significant difference in image quality was revealed between the two groups, suggesting that dose reduction in a certain limit did not affect PET image quality. Keywords: PET, 18F-FDG, BMI 1. Introduction* applied the lower and upper thresholds in the recommended range to calculate doses for whole body PET imaging in randomly selected patients. We hypothesized that dose reduction did not affect the image quality, manifested through the signal-to-noise ratio (SNR) of images. Positron Emission Tomography (PET) has been widely .