Tham khảo tài liệu 'the gauge block handbook episode 6', kỹ thuật - công nghệ, cơ khí - chế tạo máy phục vụ nhu cầu học tập, nghiên cứu và làm việc hiệu quả | Example of Calibration Output Using the 12 4 Design 1 GAUGE PROGRAM Tuesday September 10 1991 11 0 1 28 2 3 Calibration Inch Gauge Blocks Observer TS 4 Block Id s S 4361 C 1UNG X XXXXXXX Y YYYYY Federal F4 5 6 Observations Observations 7 Blocks at Ambient Conditions at Standard Conditions Differences Residuals 8 9 S C 10 Y S .22 11 X Y 12 C S 13 C X 14 Y C .89 .55 15 S X .62 16 C Y .22 17 S Y 18 X C 19 X S 20 Y X .19 21 22 Obs Within . .68 Obs Control F-test 23 Acc Within . .26 Acc Control T-test 24 Acc Group . .36 Temperature 25 26 Deviation 27 Serial Nominal Size from Nominal Total Uncertainty Coef. Block 28 Number inches microinch microinch ppm C Material 29 30 4361 steel 31 1UN6 chrome carbide 32 MARTMA steel 33 TELEDY chrome carbide 34 35 Process not in statistical control . Comments referenced by line numbers 1. The header shows which computer program was used GAGE the date and time of the calibration. The date is recorded in the MAP file. 3. Block size and operator are identified and later recorded in the MAP file. 4. Block IDs are given here. The S and C blocks are NIST master blocks. The X and Y blocks are shortened forms of customer company names represented here by strings of Xs and Ys. The ID of the comparator used in the calibration F4 is also given and recorded in 85 the MAP file. 9-20. These lines provide the comparison data at ambient temperature and corrected to 20 C using the thermal expansion coefficients in lines 30-34